  • [Journal of Magnetics.] Energy Spectrum of Electromagnetic Radiation for a Dual-Energy Cone-Beam Computed Tomography in Radiotherapy

    [Journal of Magnetics.] Energy Spectrum of Electromagnetic Radiation for a Dual-Energy Cone-Beam Computed Tomography in Radiotherapy이중 에너지 콘빔 컴퓨터 단층 촬영을 위한 전자기 방사선의 에너지 스펙트럼 연구

    울산의대, 연세의대 / 백승협, 안우상*, 안소현*

  • 출처
    Journal of Magnetics.
  • 등재일
    2020 Dec 31
  • 저널이슈번호
    Volume 25, Number 4, Pages 614-619
  • 내용

    바로가기  >

    이달의 방사선의학 연구자 바로가기 Click!

     

    Abstract

    To implement dual-energy cone-beam computed tomography (CBCT), a 1 to 3 mm thick copper filter was
    applied in front of the kilovolt generator mounted on a medical linear accelerator. The energy calibration was
    performed from three energy peaks among the electromagnetic radiation, gamma rays, emitted by the 192Ir
    source. The energy spectrum was measured using a cadmium zinc telluride (CZT) detector. A figure-of-merit
    (FOM) value was defined and evaluated in order to consider both the increase in the mean energy and the
    decrease in the intensity according to the filter. For a 1 mm thick filter, the FOM values were 0.82 and 0.92 for
    tube voltages of 100 and 120 kVp, respectively. The intens ity was maintained reasonably and the effect of
    increasing the mean energy was evident, thus indicating its suitability for implementing dual-energy CBCT.

     

    치료용 선형가속기의 kV CBCT에 대한 관전압 별, 구리 필터 두께에 따른 energy spectrum 측정 실험

     

    CZT 검출기로 측정한 Ir-192의 energy spectrum

     

     

    Affiliations

    Seunghyeop Baek (Department of Radiological Science, Yonsei University), Jin Sung Kim (Department of Radiation Oncology, Yonsei University College of Medicine,), Woo Sang Ahn * (Department of Radiation Oncology, Gangneung Asan Hospital, University of Ulsan College of Medicin), Sohyun Ahn * (Department of Radiation Oncology, Yonsei University College of Medicine)

  • 키워드
    electromagnetic spectrum; CZT detector; dual-energy CBCT; energy modulation filter; energy in electromagnetic radiation
  • 연구소개
    Dual energy CBCT를 구현하는 기존의 방법들은 X-ray 선원 또는 detector가 2개 필요하는 등 비용적인 문제가 있습니다. 이 연구에서는 X-ray generator에 energy modulation filter를 적용하여 dual energy를 구현하는 방법을 제시하고, spectrometer를 이용하여 측정한 에너지 스펙트럼의 mean energy 및 intensity를 정량적으로 분석하여 적용 가능성을 확인하고자 하였습니다.