  • [PLoS One.] Sensitivity of radio-photoluminescence glass dosimeters to accumulated doses

    국립암센터 / 김동욱, 신동호*

  • 출처
    PLoS One.
  • 등재일
    2020 Dec 3
  • 저널이슈번호
    15(12):e0234829. doi: 10.1371/journal.pone.0234829. eCollection 2020.
  • 내용

    바로가기  >

    Abstract
    Background: This study investigated the effect of accumulated doses on radio-photoluminescence glass dosimeters (RPLGDs) from measurements involving mega-voltage photons.

    Methods: Forty-five commercially available RPLGDs were irradiated to estimate their dose responses. Photon beams of 6, 10, and 15 MV were irradiated onto the RPLGDs inside a phantom, which were divided into five groups with different doses and energies. Groups 1 and 2 were irradiated at 1, 5, 10, 50, and 100 Gy in a sequential manner; Group 3 was irradiated 10 times with a dose of 10 Gy; and Groups 4 and 5 followed the same method as that of Group 3, but with doses of 50 Gy and 100 Gy, respectively. Each device was subjected to a measurement reading procedure each time irradiation.

    Results: For the annealed Group 1, RPLGD exhibited a linearity response with variance within 5%. For the non-annealed Group 2, readings demonstrated hyperlinearity at 6 MV and 10 MV, and linearity at 15 MV. Following the 100 Gy irradiation, the readings for Group 2 were 118.7 ± 1.9%, 112.2 ± 2.7%, and 101.5 ± 2.3% at 6, 10, and 15 MV, respectively. For Groups 3, 4, and 5, the responsiveness of the RPLGDs gradually decreased as the number of repeated irradiations increased. The percentage readings for the 10th beam irradiation with respect to the readings for the primary beam irradiation were 84.6 ± 1.9%, 87.5 ± 2.4%, and 93.0 ± 3.0% at 6 MV, 10 MV, and 15 MV, respectively.

    Conclusions: The non-annealed RPLGD response to dose was hyperlinear for the 6 MV and 10 MV photon beams but not for the 15 MV photon beam. Additionally, the annealed RPLGD exhibited a fading phenomenon when the measurement was repeated several times and demonstrated a relatively large fading effect at low energies than at high energies.

     

     

    Affiliations

    Dong Wook Kim  1 , Jiwon Sung  2 , Jaeman Son  2 , Han-Back Shin  1 , Min-Joo Kim  1 , Yu-Yun Noh  1 , Hojae Kim  1 , Min Cheol Han  1 , Jihun Kim  1 , Su Chul Han  1 , Kyung Hwan Chang  1 , Hojin Kim  1 , Kwangwoo Park  1 , Myonggeun Yoon  3 , Jinsung Kim  1 , Dongho Shin  4
    1 Department of Radiation Oncology, Yonsei Cancer Center, Yonsei University College of Medicine, Seoul, Korea.
    2 Department of Radiation Oncology, Seoul National University Hospital, Seoul, Korea.
    3 Department of Bio-Convergence Engineering, Korea University, Seoul, Korea.
    4 Proton Therapy Center, National Cancer Center, Goyang, Korea.

  • 편집위원

    glass dosimeter는 안정성이 좋아서 TLD를 대체하는 선량계로 사용이 확대되는 중인데 annealing 없이 반복 측정하는 경우 에너지에 따라 반응이 다름을 확인한 연구로 유리선량계를 사용하는 분들에게 유용한 정보인 것 같음