    [IEEE Trans Med Imaging.] Cardiac Image Reconstruction via Nonlinear Motion Correction Based on Partial Angle Reconstructed Images.

    KAIST/ 김승언, 나종범*

  • 출처
    IEEE Trans Med Imaging.
  • 등재일
    2017 May
  • 저널이슈번호
    36(5):1151-1161. doi: 10.1109/TMI.2017.2654508. Epub 2017 Jan 17.
  • 내용

    바로가기  >

    Abstract

    Even though the X-ray Computed Tomography (CT) scan is considered suitable for fast imaging, motion-artifact-free cardiac imaging is still an important issue, because the gantry rotation speed is not fast enough compared with the heart motion. To obtain a heart image with less motion artifacts, a motion estimation (ME) and motion compensation (MC) approach is usually adopted. In this paper, we propose an ME/MC algorithm that can estimate a nonlinear heart motion model from a sinogram with a rotation angle of less than 360°. In this algorithm, we first assume the heart motion to be nonrigid but linear, and thereby estimate an initial 4-D motion vector field (MVF) during a half rotation by using conjugate partial angle reconstructed images, as in our previous ME/MC algorithm. We then refine the MVF to determine a more accurate nonlinear MVF by maximizing the information potential of a motion-compensated image. Finally, MC is performed by incorporating the determined MVF into the image reconstruction process, and a time-resolved heart image is obtained. By using a numerical phantom, a physical cardiac phantom, and an animal data set, we demonstrate that the proposed algorithm can noticeably improve the image quality by reducing motion artifacts throughout the image. 

     

    Author information

    Kim S, Chang Y, Ra JB. 