[Appl Radiat Isot.] Development of a high-sensitivity radon monitor for a radon calibration system at KRISS 라돈 검출기 교정용 고감도 라돈모니터 개발 연구한국표준과학연구원 / 황상훈*
- 출처
- Appl Radiat Isot.
- 등재일
- 2020 Oct
- 저널이슈번호
- 164:109228. doi: 10.1016/j.apradiso.2020.109228. Epub 2020 Jun 10.
- 내용
Abstract
A radon calibration system is being developed with a high-sensitivity radon detector using electrostatic collection and an alpha spectrometer. Eight radon cells, each with a volume of 1.354 L, are integrated to achieve a high-sensitivity radon monitor with a total detector volume of 10.832 L. The sensitivity is estimated to be 35.6 ± 0.2 (36.8 ± 0.2) times higher relative to the RAD7 radon monitor for the detection of 218Po (214Po), respectively.Affiliations
S H Hwang 1 , M J Han 2 , Y G Seon 3 , J M Lee 2 , K B Lee 2
1 Korea Research Institute of Standards and Science (KRISS), Daejeon, 34113, Republic of Korea. Electronic address: shhwang@kriss.re.kr.
2 Korea Research Institute of Standards and Science (KRISS), Daejeon, 34113, Republic of Korea; University of Science & Technology (UST), Daejeon, 34113, Republic of Korea.
3 Korea Research Institute of Standards and Science (KRISS), Daejeon, 34113, Republic of Korea; Department of Physics, Kyungpook National University, Daegu, 41566, Republic of Korea.
- 키워드
- Electrostatic radon detector; Radon ((222)Rn); Radon calibration.