  • [J Appl Clin Med Phys .] Resolving signal drift in the wall-mounted camera of the RGSC system

    2024년 05월호
    [J Appl Clin Med Phys .] Resolving signal drift in the wall-mounted camera of the RGSC system

    Memorial Sloan Kettering Cancer Center / 박정훈*

  • 출처
    J Appl Clin Med Phys .
  • 등재일
    2024 Mar
  • 저널이슈번호
    25(3):e14291. doi: 10.1002/acm2.14291. Epub 2024 Feb 2.
  • 내용

    바로가기  >

    Abstract
    Purpose: To present a modified calibration method to reduce signal drift due to table sagging in Respiratory Gating for Scanner (RGSC) systems with a wall-mounted camera.

    Materials and methods: Approximately 70 kg of solid water phantoms were evenly distributed on the CT couch, mimicking the patient's weight. New calibration measurements were performed at 9 points at the combination of three lateral positions, the CT isocenter and ±10 cm laterally from the isocenter, and three longitudinal locations, the CT isocenter and ±30 cm or ±40 cm from the isocenter. The new calibration was tested in two hospitals.

    Results: Implementing the new weighed calibration method at the extended distance yielded improved results during the DIBH scan, reducing the drift to within 1 from 3 mm. The extended calibration positions exhibited similarly reduced drift in both hospitals, reinforcing the method's robustness and its potential applicability across all centers.

    Conclusion: This proposed solution aims to minimize the systematic error in radiation delivery for patients undergoing motion management with wall-mounted camera RGSC systems, especially in conjunction with a bariatric CT couchtop.

     

     

    Affiliations

    Jeonghoon Park 1, Justin D Gagneur 2, Suzanne J Chungbin 2, Yi Rong 2, Seng Boh Lim 1, Maria F Chan 1
    1Department of Medical Physics, Memorial Sloan Kettering Cancer Center, Basking Ridge, New Jersey, USA.
    2Department of Radiation Oncology, Mayo Clinic, Phoenix, Arizona, USA.

  • 키워드
    RGSC; camera calibration; motion management.
  • 연구소개
    Varian社의 RGSC 시스템은 방사선 치료용 CT 모의촬영시 CT 영상과 환자의 호흡을 동기화시켜 흉부나 복부 치료의 정확도를 향상시키는 데 사용된다. RGSC 시스템용 적외선 감시카메라가 치료실 벽에 설치된 경우, 특정 CT 장비에서 CT 영상 촬영중 호흡 신호가 왜곡되는 현상이 발견되었다. 이는 CT 장비의 환자테이블이 환자의 무게에 의해 휘어지는 것을 카메라가 제대로 인식하지 못해서 발생하는데, 심한 경우 5 mm 이상의 오차가 발생하였다. 이 연구에서는 환자의 평균무게에 해당하는 고체팬텀을 환자테이블에 올려놓고 제조사의 표준 교정거리보다 넓은 범위에서 RGSC 카메라를 교정하는 방법을 제안하였다. 새로 고안된 방법을 이용할 경우 호흡신호의 왜곡을 2 mm 이내에서 유지할 수 있었다. 환자의 몸무게에 따라 오차가 증감할 수 있으므로 RGSC를 이용한 Deep inspiration breath-hold (DIBH) 및 4D 방사선치료시 의학물리학자, 모의치료실, 치료계획실의 주의가 요구된다.