    [Radiation Measurements] Optimal design of a CsI(Tl) crystal in a SiPM based compact radiation sensor

    KAIST / 우현준, 조규성*

  • 출처
    Radiation Measurements
  • 등재일
    November 2015
  • 저널이슈번호
    Volume 82, Pages 102–107
  • 내용

    바로가기  >

    Abstract
    In this study, a compact radiation sensor for integration into mobile applications is proposed consisting of a CsI(Tl) crystal and a silicon photomultiplier with a 3 × 3 mm2 active area. Considering the size limit of mobile devices, the optimal size of the CsI(Tl) crystal was determined so that it has the best figure of merit (the absolute detection efficiency over the relative energy resolution) in the gamma energy spectrum while satisfying the criteria for the uniformity of the angular response. The final suggested dimensions consist of a cylindrically shaped crystal with a diameter of 5 mm, a total length of 5 mm and a tapered head length of 1 mm.


    Author information

    Hyunjun Yoo a, Segyeong Joo b, Shiyoung Yang c, Gyuseong Cho a,
    a Dept. of Nuclear & Quantum Engineering, Korea Advanced Institute of Science and Technology, 291 Daehak-ro, Yuseong-gu, Daejeon 305-701, Republic of Korea
    b Department of Biomedical Engineering, Asan Medical Center, University of Ulsan College of Medicine, 88, Olympic-ro 43-gil, Songpa-gu, Seoul 138-736, Republic of Korea
    c Professional Graduate School of Flexible and Printable Electronics, Chonbuk National University, 567 Baekje-dero, Deokjin-gu, Jeonju-si, Jeonbuk 561-756, Republic of Korea

  • 키워드
    CsI(Tl) crystal; Silicon photomultiplier; Absolute detection efficiency; Relative energy resolution; Angular response; Figure of merit