부산의대 / 전호상, 주지현*
Abstract
Background/aim: This study aimed to compare the dosimetric consequences of respiratory movement in volumetric-modulated arc therapy (VMAT) and three-dimensional conformal radiation therapy (3D-CRT) during postmastectomy radiation therapy, including internal mammary nodes (IMNs).
Materials and methods: Respiratory motion was implemented to a phantom using a dynamic device. The plans were delivered during cranial-caudal and ventral-dorsal movement in 5-mm (R05) and 10-mm (R10) amplitudes.
Results: At the IMN, the dose errors were -2.8% (R05) and -6.2% (R10) for 3D-CRT and -4.9% (R05) and -8.5% (R10) for VMAT. The dose errors in chest wall were -.5% (R05) and -6.0% (R10) for 3D-CRT and -1.9% (R05) and -5.3% (R10) for VMAT. The left anterior descending doses showed significantly small absolute values. The gamma pass rates of VMAT were higher than those of 3D-CRT.
Conclusions: The benefit of VMAT technique in dose distribution was maintained, except in occasional instances of large breathing motion.
Affiliations
Hosang Jeon 1 , Yongkan Ki 2 3 , Dong Woon Kim 3 , Wontaek Kim 2 4 , Jiho Nam 4 , Donghyun Kim 2 4 , Dahl Park 4 , Jongmoo Park 5 , Ji Hyeon Joo 2 3
1 Department of Radiation Oncology and Research Institute for Convergence of Biomedical Science and Technology, Pusan National University Yangsan Hospital, Yangsan, South Korea.
2 Department of Radiation Oncology, Pusan National University School of Medicine, Yangsan, South Korea.
3 Department of Radiation Oncology, Pusan National University Yangsan Hospital, Yangsan, South Korea.
4 Department of Radiation Oncology, Pusan National University Hospital, Busan, South Korea.
5 Department of Radiation Oncology, Kyungpook National University Chilgok Hospital, School of Medicine, Kyungpook National University, Daegu, South Korea.