  • [Radiat Prot Dosimetry.] COMPARISON OF SURFACE DOSES FROM INDIRECT AND DIRECT DIGITAL RADIOGRAPHY USING OPTICALLY STIMULATED LUMINESCENCE DOSEMETERS.

    순천향대 / 한수철, 장서구*

  • 출처
    Radiat Prot Dosimetry.
  • 등재일
    2018 Feb 1
  • 저널이슈번호
    178(3):322-328. doi: 10.1093/rpd/ncx113.
  • 내용

    바로가기  >

    Abstract
    An optically stimulated luminescence dosemeter was used to compare the surface dose to both eyes from an X-ray delivered frontally to the skull, and the dose could be reduced depending on image acquisition. The detectors were analysed in advance according to each image acquisition method, and the irradiation condition (mA) was obtained to equate the detective quantum efficiency of the two detectors. The surface doses to both eyes were measured in a human phantom. In the detector using the direct conversion method, the surface doses to both eyes were 0.29 ± 0.01 mSv (Rt. eye) and 0.28 ± 0.01 mSv (Lt. eye). In the detector using the indirect conversion method, the surface doses to both eyes were 0.23 ± 0.01 mSv (Rt. eye) and 0.23 ± 0.01 mSv (Lt. eye). Dose reduction by 18.00 ± 8.9% was permitted by the indirect method as compared with the direct method.

     


    Author information

    Han SC1, Na YJ2, Choi JS2, Park S1, Kang S3, Chang SG4.