  • [Med Phys.] Fully automatic quantification of transient severe respiratory motion artifact of gadoxetate disodium-enhanced MRI during arterial phase

    연세대 / 장진성, 황도식*

  • 출처
    Med Phys.
  • 등재일
    2022 Nov
  • 저널이슈번호
    49(11):7247-7261. doi: 10.1002/mp.15831. Epub 2022 Jul 10.
  • 내용

    바로가기  >

    Abstract
    Purpose: It is important to fully automate the evaluation of gadoxetate disodium-enhanced arterial phase images because the efficient quantification of transient severe motion artifacts can be used in a variety of applications. Our study proposes a fully automatic evaluation method of motion artifacts during the arterial phase of gadoxetate disodium-enhanced MR imaging.

    Methods: The proposed method was based on the construction of quality-aware features to represent the motion artifact using MR image statistics and multidirectional filtered coefficients. Using the quality-aware features, the method calculated quantitative quality scores of gadoxetate disodium-enhanced images fully automatically. The performance of our proposed method, as well as two other methods, was acquired by correlating scores against subjective scores from radiologists based on the 5-point scale and binary evaluation. The subjective scores evaluated by two radiologists were severity scores of motion artifacts in the evaluation set on a scale of 1 (no motion artifacts) to 5 (severe motion artifacts).

    Results: Pearson's linear correlation coefficient (PLCC) and Spearman's rank-ordered correlation coefficient (SROCC) values of our proposed method against the subjective scores were 0.9036 and 0.9057, respectively, whereas the PLCC values of two other methods were 0.6525 and 0.8243, and the SROCC values were 0.6070 and 0.8348. Also, in terms of binary quantification of transient severe respiratory motion, the proposed method achieved 0.9310 sensitivity, 0.9048 specificity, and 0.9200 accuracy, whereas the other two methods achieved 0.7586, 0.8996 sensitivities, 0.8098, 0.8905 specificities, and 0.9200, 0.9048 accuracies CONCLUSIONS: This study demonstrated the high performance of the proposed automatic quantification method in evaluating transient severe motion artifacts in arterial phase images.

     

     

    Affiliations

    Jinseong Jang 1, Yong Eun Chung 2, Sungwon Kim 2, Dosik Hwang 1 3 4 5
    1School of Electrical and Electronic Engineering, Yonsei University, Seoul, Republic of Korea.
    2Department of Radiology, Yonsei University College of Medicine, Yonsei University, Seoul, Republic of Korea.
    3Department of Radiology and Center for Clinical Imaging Data Science (CCIDS), Yonsei University College of Medicine, Seoul, Republic of Korea.
    4Department of Oral and Maxillofacial Radiology, Yonsei University College of Dentistry, Seoul, Republic of Korea.
    5Center for Healthcare Robotics, Korea Institute of Science and Technology, Seoul, Republic of Korea.

  • 키워드
    gadoxetate disodium; image quality assessment; motion artifact; transient severe respiratory motion.