    [J Appl Clin Med Phys.] Fabrication and evaluation of bilateral Helmholtz radiofrequency coil for thermo-stable breast image with reduced artifacts

    연세의대 / 이영한, 송규호, 김민정*

  • 출처
    J Appl Clin Med Phys.
  • 등재일
    2022 Jan
  • 저널이슈번호
    23(1):e13483. doi: 10.1002/acm2.13483. Epub 2021 Dec 1.
  • 내용

    바로가기  >

    Abstract
    Purpose: The positron emission tomography (PET)-magnetic resonance (MR) system is a newly emerging technique that yields hybrid images with high-resolution anatomical and metabolic information. With PET-MR imaging, a definitive diagnosis of breast abnormalities will be possible with high spatial accuracy and images will be acquired for the optimal fusion of anatomic locations. Therefore, we propose a PET-compatible two-channel breast MR coil with minimal disturbance to image acquisition which can be used for simultaneous PET-MR imaging in patients with breast cancer.

    Materials and methods: For coil design and construction, the conductor loops of the Helmholtz coil were tuned, matched, and subdivided with nonmagnetic components. Element values were optimized with an electromagnetic field simulation. Images were acquired on a GE 600 PET-computed tomography (CT) and GE 3.0 T MR system. For this study, we used the T1-weighted image (volunteer; repetition time (TR), 694 ms; echo time (TE), 9.6 ms) and T2-weighted image (phantom; TR, 8742 ms; TE, 104 ms) with the fast spin-echo sequence.

    Results: The results of measuring image factors with the proposed radiofrequency (RF) coil and standard conventional RF coil were as follows: signal-to-noise ratio (breast; 207.7 vs. 175.2), percent image uniformity (phantom; 89.22%-91.27% vs. 94.63%-94.77%), and Hounsfield units (phantom; -4.51 vs. 2.38).

    Conclusions: Our study focused on the feasibility of proposed two-channel Helmholtz loops (by minimizing metallic components and soldering) for PET-MR imaging and found the comparable image quality to the standard conventional coil. We believe our work will help significantly to improve image quality with the development of a less metallic breast MR coil.

     

     

    Affiliations

    Young Han Lee  1 , Kyu-Ho Song  2 , Jaemoon Yang  1 , Won Jun Kang  3 , Keum Sil Lee  4 , Min Jung Kim  3 , Eun-Kyung Kim  1 , Dan Heo  1 , Bo-Young Choe  5 , Jin-Suck Suh  1
    1 Department of Radiology, Severance Hospital, Research Institute of Radiological Science, Yonsei University College of Medicine, Seoul, Republic of Korea.
    2 Department of Radiology, Washington University School of Medicine, St. Louis, Missouri, USA.
    3 Department of Nuclear Medicine, Severance Hospital, Yonsei University College of Medicine, Seoul, Republic of Korea.
    4 Department of Radiology, Stanford University School of Medicine, Stanford, California, USA.
    5 Department of Biomedical Engineering, College of Medicine, and Research Institute of Biomedical Engineering, The Catholic University of Korea, Seoul, Republic of Korea.

     

  • 키워드
    breast imaging; magnetic resonance imaging; metallic artifact; positron emission tomography; radiofrequency coil.
  • 덧글달기
    덧글달기
       IP : 3.128.79.107

    등록