[J Appl Clin Med Phys .] Quantitative assessment of CBCT image quality metrics and their translation to clinical Art workflow performance in breast cancer patients가톨릭의대 / 김선화, 강영남*
- 출처
- J Appl Clin Med Phys .
- 등재일
- 2026 Jun
- 저널이슈번호
- 27(6):e70648.
- 내용
Abstract
Background: Online adaptive radiotherapy (ART) relies on cone-beam computed tomography (CBCT) imaging for daily treatment adaptation, but CBCT's physical limitations can compromise contour delineation and dose calculation accuracy. The relationship between image quality metrics and clinical ART workflow accuracy remains unclear.Purpose: To investigate the clinical impact of CBCT image quality on ART workflow accuracy in breast cancer patients by comparing contour agreement and dose calculation accuracy between Halcyon and TrueBeam systems, establishing how quantitative image quality metrics affect contour agreement and dose calculation accuracy within the ART workflow.
Methods: Forty breast cancer patients treated with VMAT were equally divided between Halcyon (n = 20) and TrueBeam (n = 20). Phantom-based image quality assessment evaluated HU uniformity, contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), and modulation transfer function (MTF) using Catphan 504. Planning CT and first-day CBCT images were aligned using deformable registration. AI auto-segmentation generated contours for six structures, evaluated using Dice Similarity Coefficient (DSC) and other geometric metrics. VMAT plans were recalculated on CBCT images using system-specific HU-to-electron density calibration curves for dose-volume histogram, 3D gamma analysis, and dose difference analysis.
Results: Halcyon demonstrated 1.65-fold higher CNR (22.21 vs. 13.42) and 1.78-fold better HU uniformity (9 vs. 16 HU) compared to TrueBeam, despite comparable spatial resolution (MTF50: 0.36 lp/mm). Contour agreement analysis showed Halcyon achieved higher accuracy (treated breast DSC: 0.92 ± 0.06 vs. 0.87 ± 0.21, p < 0.001) with 3.5-fold lower inter-patient variability. Dosimetric analysis revealed paradoxical patterns: TrueBeam maintained better PTV coverage (D95% difference: 0.013 vs. 0.207 Gy) but showed significantly increased internal dose metrics (Dmean difference: -0.320 vs. 0.046 Gy, p < 0.001). Although both systems achieved > 99.8% gamma passing rates at 3%/3 mm criteria, stringent 1%/2 mm criteria revealed substantial performance differences (Halcyon: 97.66 ± 1.91% vs. TrueBeam: 90.28 ± 6.51%, p < 0.001).
Conclusion: Quantitative CBCT image quality metrics, particularly CNR and HU uniformity, directly affect both contour agreement and dosimetric concordance in breast ART workflows. These findings provide quantitative evidence linking CBCT image quality metrics to both contour agreement and dose calculation accuracy within the ART workflow.
Affiliations
Sunhwa Kim 1, Young Kyu Lee 2, Wonjoong Cheon 2, Yunji Seol 3, Byung-Ock Choi 2, Young Nam Kang 2
1Department of Medical Sciences, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea.
2Department of Radiation Oncology, College of Medicine, Seoul St. Mary's Hospital, The Catholic University of Korea, Seoul, Republic of Korea.
3Department of Proton Center, National Cancer Center of Korea, Goyang-si, Gyeonggi-do, Republic of Korea.
- 키워드
- CBCT image quality; adaptive radiotherapy; breast cancer; contour agreement; dosimetric accuracy.
- 덧글달기
- 이전글 [J Appl Clin Med Phys .] The illusory correlation between geometrical similarity and physician satisfaction in heart auto-segmentation revealed by a physician-blind test
- 다음글 [Phys Med Biol .] A proof-of-concept automated method for accurate skin dosimetry: correcting overestimated surface dose measurements






편집위원
- 본 논문은 CBCT 영상 기반의 온라인 맞춤형 방사선치료(ART)의 정확성에 관한 연구로서, 두 종류의 CBCT에 대하여 화질 지표와 선량분포의 정확성을 정량적으로 분석하였으며, 정확한 치료계획을 위해 CBCT 영상의 CNR과 SNR과 같은 화질 지표들이 중요함을 강조하고 있다.
- 연구자들은 Catphan-504 팬텀을 이용하여 Halcyon CBCT와 TrueBeam CBCT의 화질을 비교한 결과 Halcyon CBCT가 CNR과 SNR에 있어서 높은 화질을 보였으며, 다수의 유방암 환자의 CBCT 영상을 이용한 선량분포 계산결과에 있어서도 Halcyon CBCT가 보다 우수함을 보였다.
※ ART: Adaptive radiotherapy
※ CNR: Contrast-to-noise ratio
※ SNR: Signal-to-noise ratio