  • [Med Phys.] Feasibility of isodose-shaped scintillation detectors for the measurement of gamma knife output factors 감마나이프 출력 계수 측정을 위한 등선량 형태의 신틸레이션 계측기 개발 연구

    한양대, 서울의대 / 김태훈, 김용균*, 정현태*

  • 출처
    Med Phys.
  • 등재일
    2022 Mar
  • 저널이슈번호
    49(3):1944-1954. doi: 10.1002/mp.15469. Epub 2022 Jan 28.
  • 내용

    바로가기  >

    Abstract
    Purpose: Scintillation detectors were 3D printed based on a gamma knife (GK) dose distribution to calculate the volume averaging effect. The collimator output factors were measured using isodose-shaped scintillators (ISSs) and compared with those of a micro-diamond detector and previous reports.

    Methods: An absorbed dose distribution in a spherical dosimetry phantom with a radius of 8 cm was obtained from GK treatment planning software (Leksell GammaPlan [LGP], Elekta AB, Stockholm, Sweden). Two types of ISSs were fabricated to fit the 97.2% (ISS-1) and 95.6% (ISS-2) isodose surfaces. The volume averaging correction factors were obtained by dividing the absorbed dose to water in the central voxel (CV) by that in the ISS. The correction effect due to the difference between the ISS and water was calculated by Monte Carlo simulations. Ten ISS detectors, five of each type, were used to measure the output factors of the 4- and 8-mm collimators of a GK Icon to assess system consistency. The output factors of seven GKs were measured using two ISS detectors, one of each type, and a PTW T60019 (PTW, Freiburg, Germany) micro-diamond detector.

    Results: The detector output ratios (DORs) measured using the five ISSs of each type were consistent, with standard uncertainties less than 0.2%. In the 4-mm field, the volume averaging correction factor ratios were 1.018 and 1.026, and the output factors after all corrections were 0.827 (0.006) and 0.825 (0.006) for ISS-1 and ISS-2, respectively. In the 8-mm field, the volume averaging correction factor ratios were 1.000 for both ISS types, and the output factors were 0.898 (0.003) and 0.900 (0.003) for ISS-1 and ISS-2, respectively. The ISS detectors could measure the output factors of a GK with uncertainties comparable to that of the PTW 60019 detector. The output factors of all detectors decreased with the dose rate.

    Conclusion: The volume averaging effect of an ISS developed in-house could be calculated using known dose distributions. The collimator output factors of the GK Perfexion/Icon models measured using ISS detectors were consistent with those of a commercial synthetic micro-diamond detector and recent studies.

     

     

    Affiliations

    Tae Hoon Kim  1 , Hye Jeong Yang  2 , Jae Young Jeong  1 , Thomas Schaarschmidt  1 , Yong Kyun Kim  1 , Hyun-Tai Chung  3
    1 Department of Nuclear Engineering, Hanyang University College of Engineering, Seoul, Republic of Korea.
    2 Department of Biomedical Engineering, College of Medicine, Catholic University of Korea, Seoul, Republic of Korea.
    3 Department of Neurosurgery, Seoul National University College of Medicine, Seoul, Republic of Korea.

  • 키워드
    3D printed scintillator; gamma knife; output factor; scintillation dosimeter; small field dosimetry.
  • 편집위원

    상용화된 단일 복셀 검출기는 출력 불확실성이 1%를 초과하기 때문에 비실용적이나, 논문에서 개발한 검출기는 0.2% 미만의 표준 불확실성으로 측정되어 선량 계산에 높은 정확도를 얻을 수 있습니다.
    3D 프린팅기술을 이용하여 다양한 형태의 검출기에 적용이 가능할 뿐만 아니라, 검출기의 등선량 표면과 일치하도록 선택하여, 섬광 검출기의 부피 평균 효과를 계산이 가능하도록 고려하였기에 다양한 응용이 가능하도록 고려되었습니다.
    개발한 isodose-shaped scintillator 검출기를 사용하여 측정된 감마 나이프 Perfexion/Icon 모델의 콜리메이터 출력 계수가 상용 합성 마이크로 다이아몬드 검출기 및 최근 연구의 결과와 잘 일치하는 우수한 결과를 얻음으로써, 국내 검출기 연구의 우수성이 돋보입니다.