  • [Phys Eng Sci Med.] Assessment of dosimetric leaf gap correction factor in Mobius3D commissioning affected by couch top

    [Phys Eng Sci Med.] Assessment of dosimetric leaf gap correction factor in Mobius3D commissioning affected by couch top 치료테이블이 다엽콜리메이터의 선량학적 엽 간격에 미치는 영향 연구

    연세대 / 한민철, 홍채선*, 김진성*

  • 출처
    Phys Eng Sci Med.
  • 등재일
    2020 Jul 22.
  • 저널이슈번호
    doi: 10.1007/s13246-020-00905-z. O;
  • 내용

    바로가기  >

    이달의 방사선의학 연구자 바로가기 Click!

     

    Abstract
    This study assesses the dosimetric leaf gap (DLG) correction factor in Mobius3D commissioning affected by a couch top platform and calculates the optimal DLG value according to the point dose difference function. DLG optimizations were performed for 3 LINAC machines and a total of 30 patient volumetric modulated arc therapy plans (i.e., 10 plans per each LINAC). Point dose calculations were performed using an automatic dose calculation system in Mobius3D as well as Mobis3D calculation using a Mobius Verification Phantom (MVP)-based quality assurance plan with a carbon fiber couch top. Subsequently, the results were compared with measurement data. The averaged point dose measured for the MVP with a couch top decreased by approximately 2% relative to that without the couch top. The average of the optimal DLG factors increased by 1.153 mm due to the couch top effect for a dose decrease of 2% at the measured point. In the procedure of Mobius beam commissioning, users should adjust the DLG correction factor using a specific phantom (including MVP) with a couch top structure. If the DLG optimization were performed by using MVP automatic dose calculation system, the factor should be increased by approximately 1.2 mm per 2% dose difference considering user's couch top effect.

     

    본 그림은 Mobius3D에서 commissioning 및 QA 용으로 제작된 전용 팬텀(MVP)에 본원에서 사용하는 Couch top을 추가하여 선량계산된 모습을 보여줍니다.

     

     

    Affiliations

    Min Cheol Han  1 , Jihun Kim  1 , Chae-Seon Hong  2 , Kyung Hwan Chang  1 , Su Chul Han  1 , Kwangwoo Park  1 , Dong Wook Kim  1 , Myoung Kyu Park  3 , Yu Yun Noh  3 , Jin Sung Kim  4
    1 Department of Radiation Oncology, Yonsei University College of Medicine, 50-1 Yonsei-ro, Seodaemun-gu, Seoul, South Korea.
    2 Department of Radiation Oncology, Yonsei University College of Medicine, 50-1 Yonsei-ro, Seodaemun-gu, Seoul, South Korea. cs.hong@yuhs.ac.
    3 Department of Radiation Oncology, Yonsei Cancer Center, Seoul, South Korea.
    4 Department of Radiation Oncology, Yonsei University College of Medicine, 50-1 Yonsei-ro, Seodaemun-gu, Seoul, South Korea. jinsung@yuhs.ac.

  • 키워드
    Couch effect; Dosimetric leaf gap; Mobius; Mobius verification phantom.
  • 연구소개
    환자 맞춤형 선량계산 시스템으로 활용되는 Mobius3D를 활용하기 위해서는 각 LINAC 장비마다 가지는 고유의 Dosimetric leaf gap (DLG) 차이를 반영해주어야 합니다. 이를 평가하기 위해 프로그램 내에서는 Mobius3D 전용 팬텀(MVP) 및 이를 기반으로한 자동 선량계산 시스템을 제공하지만, 이는 치료장비의 Couch top을 고려하지 않고 계산되기에 실제 측정값과 차이가 발생하게 됩니다. 본 연구는 Mobius3D 내에서 Couch top에 의한 선량차이가 DLG correction factor에 미치는 영향을 평가하였습니다.
  • 편집위원

    근래의 방사선치료 환자정도관리로 epid 측정값이나 MLC position의 log 기록을 이용하여 간접적인 계산방식으로 수행하는 경우가 증가하고 있고, 이에 따라 관련 소프트웨어를 도입하는 병원도 증가하고 있는데, 이러한 소프트웨어중의 하나인 Mobius3D를 사용할 때 고려해야 할 부분을 잘 분석하고 대안을 제시한 것 같음