방사선방호 및 안전

본문글자크기
  • [Journal of Radioanalytical and Nuclear Chemistry] Efficiency enhanced electrolysis-based tritium continuous monitor

    UNIST / 배준우, 김희령*

  • 출처
    Journal of Radioanalytical and Nuclear Chemistry
  • 등재일
    2019 December
  • 저널이슈번호
    Volume 322, Issue 3, pp 1323–1329
  • 내용

    바로가기  >

    Abstract
    Developing an online monitoring system becomes useful for continuously detection of tritium in water. A plastic scintillator chamber was used to tritium detection in hydrogen gas produced by a proton exchange membrane electrolysis cell. The fractionation factor (proton selectivity) was estimated as being of 5.41 ± 0.20, and detection efficiency of the plastic scintillator chamber was 4.25 ± 0.22%. For liquid samples, the minimum detectable activity of the developed system was estimated as 655 kBq L−1. Using this online technique for tritium monitoring based on electrolysis is expected to be useful for analysis of large tritiated water bearing samples.

  • 키워드
    Proton exchange membrane; Electrolysis; Tritium; Continuous monitoring; Fractionation factor
  • 덧글달기
    덧글달기
       IP : 3.147.89.85

    등록