  • [Med Phys.] Systematic study on factors influencing the performance of interdetector scatter recovery in small-animal PET.

    서울대 / 이승은, 이재성*

  • 출처
    Med Phys.
  • 등재일
    2018 May 31. doi: 10.1002/mp.13020. [Epub ahead of
  • 저널이슈번호
  • 내용

    바로가기  >

    Abstract
    PURPOSE:
    Interdetector scatter (IDS) is a triple coincidence caused by the Compton scatter of an annihilation photon from one detector block to another which frequently occurs in small-animal positron emission tomography (PET). By finding the true lines-of-response (LORs) of annihilation photon pairs among three possible LORs in IDS events, we can utilize these recovered events to improve the sensitivity of PET systems. IDS recovery should be accurate to yield reliable images with relatively short scan times. We systemically investigated physical factors affecting IDS recovery performance, focusing on the reconstructed image quality of small-animal PET. We evaluated sensitivity increase, recovery accuracy, and image quality by applying different combinations of energy window, recovery scheme, and scanner properties.

    METHODS:
    We used GATE Monte Carlo simulation to acquire coincidence events from a NEMA NU 4-2008 image quality phantom using small-animal PET scanner with axial field of view of 55 mm and diameter of 64 mm. We first defined energy window criteria to obtain valid IDS events. Their role was to assign triple coincidences as IDS events and to restrict the number of LOR candidates to two. We tested three different energy windows around 511 keV. Second, we applied four different recovery schemes (maximum energy, Compton kinematics, neural network, and proportional) to assigned IDS events. To measure the effects of scanner properties, energy resolutions of 0-20% and one to four depth-of-interaction (DOI) layers were simulated. For every combination of the factors, we measured sensitivity increase and recovery accuracy. We also analyzed the reconstructed images for each IDS recovery method in terms of mean pixel intensity, noise, signal-to-noise ratio (SNR), contrast, and recovery coefficients.

    RESULTS:
    Sensitivity increase depended on the energy window and energy resolution. The maximum increase in sensitivity was 33% when energy window of [250, 750] keV was applied. Higher energy resolution yielded larger sensitivity increase, especially for narrow windows. Recovery accuracy was affected by all the factors tested in this study. Accuracy increased with narrower energy window, and a neural network scheme was the most accurate. The better energy resolution and DOI capability improved accuracy by providing precise measurement of energies and interaction positions. In image quality analysis, noise and SNR were highly dependent on the sensitivity increase and energy window. When the same energy window was applied, SNR, contrast, and recovery coefficients were higher with higher accuracy of the scheme. Meanwhile, the proportional scheme yielded the best image quality among the schemes and reduced 20% of scan time to achieve the same SNR as that of double coincidence images.

    CONCLUSIONS:
    As a fundamental research for real implementation of IDS recovery, we conducted a simulation study to evaluate the factors affecting sensitivity increase, recovery accuracy, and image quality. Sensitivity increase was dependent on the energy window and energy resolution, while the recovery accuracy was affected by energy window, recovery scheme, energy resolution, and DOI capability. In image quality analysis, sensitivity increase and recovery accuracy dominantly affected the noise and quantitative accuracy, respectively. Among the recovery schemes, the proportional scheme obtained the best image quality.

     


    Author information

    Lee S1,2, Lee MS2,3, Kim KY1,2, Lee JS1,2,3,4.
    1
    Department of Biomedical Sciences, Seoul National University College of Medicine, Seoul, 03080, Korea.
    2
    Department of Nuclear Medicine, Seoul National University College of Medicine, Seoul, 03080, Korea.
    3
    Interdisciplinary Program in Radiation Applied Life Science, Seoul National University, Seoul, 03080, Korea.
    4
    Institute of Radiation Medicine, Medical Research Center, Seoul National University, Seoul, 03080, Korea.

  • 키워드
    interdetector scatter
  • 편집위원

    작은 동물용 PET에서 IDS에 기인하는 triple coincidence에 의한 Sensitivity와 Accuracy를 energy window, energy resolution등에 따라 분석하였다. PET은 아직도 개척할 요소가 많이 있음을 새삼 느꼈다.