  • [Med Phys.] Surface applicator of a miniature X-ray tube for superficial electronic brachytherapy of skin cancer.

    KAIST/ 김현남, 조성오*

  • 출처
    Med Phys.
  • 등재일
    2017 Nov 6.
  • 저널이슈번호
    doi: 10.1002/mp.12656. [Epub ahead of ;
  • 내용

    바로가기  >

    Abstract
    PURPOSE:
    We designed and fabricated a surface applicator of a novel carbon nanotube (CNT)-based miniature X-ray tube for the use in superficial electronic brachytherapy of skin cancer. To investigate the effectiveness of the surface applicator, the performance of the applicator was numerically and experimentally analyzed.

    METHODS:
    The surface applicator consists of a graphite flattening filter and an X-ray shield. A Monte Carlo radiation transport code, MCNP6, was used to optimize the geometries of both the flattening filter and the shield so that X-rays are generated uniformly over the desired region. The performance of the graphite filter was compared with that of conventional aluminum (Al) filters of different geometries using the numerical simulations. After fabricating a surface applicator, the X-ray spatial distribution was measured to evaluate the performance of the applicator.

    RESULTS:
    The graphite filter shows better spatial dose uniformity and less dose distortion than Al filters. Moreover, graphite allows easy fabrication of the flattening filter due to its low X-ray attenuation property, which is particularly important for low-energy electronic brachytherapy. The applicator also shows that no further X-ray shielding is required for the application because unwanted X-rays are completely protected. As a result, highly uniform X-ray dose distribution was achieved from the miniature X-ray tube mounted with the surface applicators. The measured values of both flatness and symmetry were less than 5% and the measured penumbra values were less than 1 mm. All these values satisfy the currently accepted tolerance criteria for radiation therapy.

    CONCLUSIONS:
    The surface applicator exhibits sufficient performance capability for their application in electronic brachytherapy of skin cancers.

     


    Author information

    Kim HN1, Lee JH1, Park HB1, Kim HJ2, Cho SO1.
    1 Department of Nuclear and Quantum Engineering, Korea Advanced Institute of Science and Technology, Daejeon, 305-701, Korea.
    2 Department of Radiation Oncology, Asan Medical Center, Seoul, 05505, Korea.

  • 키워드
    electronic brachytherapy; miniature X-ray tube; skin cancer; surface applicator