  • [Radiation Measurement] Reference thermal neutron field at KRISS for calibration of neutron detectors

    한국표준과학연구원, 한양대/ 김윤호, 박현서*,김용균*

  • 출처
    Radiation Measurement
  • 등재일
    2017 Oct
  • 저널이슈번호
  • 내용

    바로가기  >

    Abstract

    A reference thermal neutron field has been established at the Korea Research Institute of Standards and Science (KRISS) by using a 241Am-Be neutron source and a high-purity graphite pile constructed by stacking graphite blocks. The properties of the graphite blocks such as impurities, density, and dimensions were studied thoroughly to understand the characteristics of the generated field. The energy spectrum and thermal neutron fractions were simulated with the Monte Carlo N-Particle eXtended code using measured physical parameters. The neutron effective temperature was 308 K, and the fraction of thermal neutrons was approximately 95% at the reference position of the thermal neutron field. The thermal neutron fluence rate was determined by adopting the Westcott convention method based on neutron activation analysis using a gold foil. The Westcott fluence rate for thermal neutron at the reference position was (2326.7 ± 8.4) cm−2s−1. The true thermal neutron fluence rate at the reference position in the KRISS thermal neutron field was (2700 ± 29) cm−2s−1 (at the reference date of June 30, 2014). The response of a spherical proportional counter with He-3 (SP9 neutron detector) was evaluated in the newly established field as (3.083 ± 0.045) cm2 for the reference calibration condition (a parallel neutron beam with a Maxwellian energy distribution having a most probable energy of 0.025 eV). 


    Yun HoKim a, HyeonseoPark a, Yong KyunKim b, JunghoKim a, JeongsooKang a

    a Korea Research Institute of Standards and Science, Daejeon 305–340, Republic of Korea

    b Department of Nuclear Engineering, Hanyang University, Seoul 133–791, Republic of Korea

     

  • 키워드
    Reference thermal neutron fieldThermal neutron fluenceNeutron activation analysisThermal neutron detector calibration
  • 편집위원

    중성자 측정기에 대한 보정을 위한 연구가 최근 관심을 끌고 있는 붕소를 이용한 중성자치료기 연구와 더불어 관심을 끕니다.